Contents
SOP (Standard Operating Procedure) of hot air oven Download Notes PDF BPharmacy 7th Semester 2022
π BP701T Instrumental Methods of Analysis
Title: |
SOP (Standard Operating Procedure) of hot air oven |
Course: | BPharmacy |
Semester/Year: | 7th Semester |
Subject: | BP701T Instrumental Methods of Analysis |
Session: | 2022 |
Category: | Notes |
BPharmacy Practical and Experiments Standard Operating Procedure (SOP) Pharmaceutical Engineering |
Β
You can download BPharmacy Notes & Previous Year’s Question Paper from here. All semester-wise Notes & Previous Year Question Papers are available on HK Technical. You can download all question papers & notes for BPharmacy all semester/years for free.
You can view All Question Papers From Here.
π Download All Question Papers for BPharmacy
π Premium Question Paper/Notes Library
Swami Dayanand post graduate institute of pharmaceutical sciences uhs,rohtak
S.O.P. OF HOT AIR OVEN
Safety Precautions
β’ Maximum Temp.: 350oC.
β’ Ensure that the Exhaust blower is ON before starting the oven.
β’ Ensure that temperature does not shoot higher than the set temperature.
β’ Always wear the gloves while handling the hot trays.
β’ Avoid touching the exhaust of the system.
β’ Before removing side cover make sure equipment is switched off.
Cleaning Methods
1. Cleaning procedures should be done only with iso-propyl alcohol using lint free cloth.
2. The SS trays should be cleaned with iso-propyl alcohol using lint-free cloth.
STANDARD OPERATIONAL PROCEDURE
Switching On Procedure –
I. Ensure that gas/air flow cylinders are ON from facility side.
II. Turn ON the Rota meter for air flow in the oven.
III. Switch ON the Mains of the Oven.
IV. Turn ON the mains. The Red LED will glow means oven is ON.
V. Switch ON the fan.
VI. Once the system is on you can set the required temperature on the Digital
Temperature controller
VII. Press βSETβ once on digital controller.
VIII. Press UP arrow β to set the required temperature. The maximum temperature is
350Β°C.
IX. Press βENTβ to start the HEATING UP program.
X. The YELLOW LED will glow indicating the heater is ON.
XI. Wait till the required temperature is reached.
Switching Off Procedure –
1) Once the process is over set the required shut down temperature on the Digital
Temperature controller as follows:
β’ Press βSETβ once
β’ Use DOWN arrow β to reduce the oven temperature to 25oC.
β’ Press βENTβ to start ramp down program.
β’ The YELLOW LED will stop glowing indicating heater is OFF.
2) Make a log book entry
π BP701T Instrumental Methods of Analysis
π BPharmacy 7th Semester Notes
π Download Previous Year’s Question Papers
π Download Notes PDF
π Download Handwritten Notes
π Download BPharmacy Question Papers
π Download B.Sc. Nursing Question Papers
π Download Last Year’s Question Papers
π BPharmacy All Semester PDF Download
π Download PDF Notes as per Latest PCI Syllabus Pattern
π Download B.Sc. Nursing Notes as per latest syllabus pattern
π Download All Previous Years Question Papers
π PGIMS Rohtak Question Papers
π AIIMS Question Papers
π B.Sc. Nursing Question Papers
π B.Tech Question Papers
π Download Free PDF Notes
Β